Product introduction

【Indication】

1. When tested with normal or corrected vision, the inner and outer surfaces of the nozzle shall be smooth and clean, and the pipe wall shall be free of ripples, gel, bubbles and impurities;

2. The protective sleeve of connecting pipe shall be fastened;

3. No leakage: the connection between the connecting pipe and the peritoneal dialysis tube shall be able to withstand the pressure of 50KPa without leakage;

4. The connecting part between the connecting pipe and the peritoneal dialysis pipe and the connecting parts of the connecting pipe shall be firmly connected and shall be able to bear the tensile force of 15N;

5. The reversing of connecting pipe shall be convenient and reliable;

6. Breaking force of nozzle: the minimum breaking force shall be greater than 3N;

7. The connecting pipe shall be sterile;

8. The connecting pipe shall be sterilized with ethylene oxide. After 15 days, the residual amount of ethylene oxide shall not be greater than 10 μ g/g;

9. The total content of barium, chromium, copper, lead and tin in the test solution shall not exceed 5 μ G / ml, cadmium content should be ≤ 0.1 μ g/mL;

10. The pH value of the leaching solution of the nozzle shall be compared with the blank control solution of the same batch, and the difference of pH value shall not exceed 1.0.
【purpose/Scope of usage】
The product is used for connecting with the external part of peritoneal dialysis tube in hospitals and families.
【Usage Guidelines/ Instruction for usage.】
1. Wear a mask and thoroughly clean your hands;

2. Before use, first check whether the packaging of the product is intact;

3. The operator shall stand on a smooth plane and grasp transfer set, tear the package at the tear opening to expose transfer set;

4. Take out transfer set from the package;

5. Take out the connecting pipe, one end is connected with the built-in pipe with a titanium alloy joint, and the other end is connected with the dialysate pipe; When replacing, grasp transfer set and rotate it counterclockwise to remove transfer set;

6. After the pipeline connection is completed, place the rotating water stop sleeve at the liquid inlet position according to the indicated direction, and confirm it again, open the liquid stop clamp;

7. After the liquid inlet is completed, put the water stop sleeve at the liquid outlet position and drain the liquid in the abdominal cavity. This cycle is completed the dialysis treatment process;

8. After dialysis treatment, connect a new minicap to the cathode locking connector. Replace the minicap after each use.
【Contraindication】
1. The product is only used for peritoneal dialysis and is prohibited for those with other diseases;

2. It is forbidden for patients with severe infection
【Warning】
1. The nozzle is disposable;

2. Do not use if the package is opened or the sponge inside the iodophor cap is dry before use;

3. Connect correctly according to the schematic diagram;

4. The minicap shall be replaced every day and shall not be reused;

5. Aseptic operation is required when using this product;

6. Please read the instructions attached to the product carefully before use.
【Caution】
1. Do not wipe the connecting parts with hydrogen peroxide, alcohol or alcohol containing anti-corrosion substances;

2. Do not allow bleach to enter the connected parts;

3. Do not touch the sponge in the Minicap with your hand during any process;

4. The equipment cannot be purchased at will. Please follow the doctor's advice.
【Manufacture Date】

【Sterilization Date】

【Product validity】The validity period is 2 years from the date of sterilization.
【Product storage conditions】
The product shall be placed in a clean, dry and well ventilated room with relative humidity no more than 80% and no corrosive gas.


You also like